SC4AJ Conductivity Sensor with Cell Constants 0.02 cm-1 and 0.1 cm-1 for Temperature Range 0 to 110℃

Cell constant 0.02 cm-1 Conductivity Sensor

Cell constant 0.1 cm-1 SC4AJ

Temperature Range 0 to 110℃ Water Quality Sensor

Marca:

Yokogawa

Numero di modello:

Sensore SC4AJ

Contattaci
Chiedi un preventivo
Dettagli del prodotto
Principio di misurazione:
Sistema a due elettrodi
Costante cellulare:
0,02 cm-1, 0,1 cm-1
Intervallo di temperatura:
Da 0 a 110 ℃ (per elettrodo)
Sterilizzazione per elettrodo:
135 ℃ (275 ℉), entro 30 minuti dalla sterilizzazione a vapore
Intervallo di pressione:
Da 0 a 1 MPa (per elettrodo)
Evidenziare:

Cell constant 0.02 cm-1 Conductivity Sensor

,

Cell constant 0.1 cm-1 SC4AJ

,

Temperature Range 0 to 110℃ Water Quality Sensor

Termini di pagamento e di spedizione
Quantità di ordine minimo
1pcs
Imballaggi particolari
Ogni unità ha una singola scatola e tutte le caselle sono imballate in pacchetti standard o richiest
Tempi di consegna
5-8 giorni di lavoro
Termini di pagamento
Moneygram, Western Union, T/T, D/P, D/A, L/C.
Capacità di alimentazione
1000 pezzi/pezzi a settimana negoziabile
Descrizione del prodotto
Conductivity Sensor SC4AJ
Yokogawa has been supplying superior on-line analyzers for monitoring or controlling the conductivity of liquid solutions. Our product line includes the FLXA402 four-wire conductivity/resistivity converter, along with the FLXA202 and FLXA21 two-wire conductivity/resistivity transmitters, complemented by a range of precision sensors for accurate liquid conductivity measurement.
The SC4AJ conductivity sensor features a compact, convenient design suitable for diverse applications including boiler water monitoring, hydroponics conductivity measurement, and ultrapure water resistivity measurement in semiconductor, pharmaceutical, food, and power industries.
Specifications
Measurement Object Conductivity of liquids or solutions
Measuring Principle Two-electrode system
Cell Constant 0.02 cm-1, 0.1 cm-1
Measuring Range
  • 0 - 0.5 to 0 - 200 μS/cm (Cell constant: 0.02 cm-1, with SC450G, SC202)
  • 0 - 100 μS/cm to 0 - 1 mS/cm (Cell constant: 0.02 cm-1, with SC100)
  • 0 - 5 μS/cm to 0 - 1 mS/cm (Cell constant: 0.1 cm-1, with SC450G, SC202)
  • 0 - 500 μS/cm to 0 - 5 mS/cm (Cell constant: 0.1 cm-1, with SC100; Material: Titanium)
Temperature Range 0 to 110℃ (For electrode)
Sterilization for Electrode 135℃ (275℉), within 30 minutes in Steam Sterilization
Pressure Range 0 to 1 MPa (For electrode)
Sample Solution Condition Although flow rate is not limited in measurement, air bubbles should not be mixed in the sample solutions to obtain correct measured values.
Temperature Sensor Pt1000
Weight Sensors:
  • Adapter mounting type (SC4AJ-S-AD-09-002-05): approx. 450 g
  • Adapter mounting type (SC4AJ-S-AD-15-002-05): approx. 520 g
  • Welding socket type (SC4AJ-S-SA-NN-002-05): approx. 670 g
  • 5" welding clamp type (SC4AJ-S-SB-NN-002-05): approx. 550 g
  • 2" welding clamp type (SC4AJ-S-SC-NN-002-05): approx. 670 g
Adapters:
  • 3/4" stainless steel adapter (/PS): approx. 110 g
  • 3/4" PVDF adapter (/PF): approx. 35 g
  • Straight welding socket (/SA1): approx. 300 g
  • Angle welding socket 15 (/SA2): approx. 320 g
  • Welding clamp 1" (/SB1): approx. 330 g
  • Welding clamp 1.5" (/SB2): approx. 305 g
  • Welding clamp 2" (/SC1): approx. 350 g
Process Connection Adapter mounting, welding socket, welding clamp